Lecture 01-Introduction

IIT Roorkee July 201
61 followers
Dec 2, 2023
Definitions and introduction of sensors and biosensors were presented and components of a sensor, their working and types were discussed in detail.
I welcome you all to the first lecture of Optical Sensors course. Today, we are going
to discuss the very basics of the sensors and biosensors and the components involved
in this and what are the basic characteristics. So, here is the outline of the talk. We will
first introduce sensors and biosensors and then we will see the market overview and
need of biosensors. Why do we need them and what are the components and their functions
and finally, we will discuss optical parameters which are required for transduction in this
kind of sensors. A sensor is a device which measures a change
in one physical parameter in terms of magnitude of a second different parameter which can
be measured more conveniently and perhaps more accurately. What is the meaning of this
thing? This means that you want to measure something in terms of something else which
can be measured more conveniently and accurately. I have listed here few examples of sensors.
For example, this thermometer, what it does? It measures temperature, but in terms of what?
In terms of the length of this column. So, you see that if you increase the temperature
of this room say by 2 degree centigrade, you hardly feel any change, ok. But this device
it can tell you how much change in temperature occurred by measuring the change in length
actually. So, it measures the length which can be measured more conveniently and more
accurately and then it translates in temperature. Another example is this old time coal miners
biosensor this yellow canary bird which actually coal miners used to bring to coal mines in
old times and what happened actually that when there was some unpleasant situation say
some leakage of methane gas or something, then it will start crying. So, they know that
now this is the time to evacuate the mines. I think that they chose a yellow bird because
they wanted to see it easily even in dark. Now, you can see a conventional glucometer
which is commercially available in the market. It actually measures current. It is an electrochemical
sensor and it translates this change in current in terms of glucose concentration in blood.
There is a conventional pregnancy kit which measures the HCG hormone in urine and this
infectious disease biosensor detects tuberculosis. Say for example, the patient has to go and cough
in here and from the vapors you know that the person has tuberculosis or not. So, these
are some kind of sensors, but out of these sensors what you can do is that we can categorize
for example, these five sensors in two groups. One group is a sensor which is quantitative
while the other sensor is purely qualitative. What does it mean? It means, say for example,
this coal miners biosensor yellow canary bird or pregnancy kit. It tells you that if
the situation is there or not. For example, this coal miners biosensor tells you that
there is an unpleasant situation, but it does not tell you that how unpleasant. It does
not tell you that how much amount of gas leakage is there, but there is something which is
unpleasant and then it starts crying. This pregnancy kit tells that either the lady
is pregnant or not. It does not give you any information that how much pregnant. So, it
is a qualitative sensor. Now, come to the thermometer or glucometer.
It tells that the patient, let us say focus to the thermometer. So, the thermometer says
how much change in temperature occurs. It says that there is a change in temperature
and also that how much change in temperature occurs. So, it is quantitative. Similarly,
a glucometer tells you that there is a change in glucose concentration in blood and how
much change. It also measures the concentration. So, it is qualitative and quantitative. So,
we can divide this whole branch of sensors in two parts. One is qualitative, others is
quantitative. So, they have different uses, you could have seen from here.
Let us see what is a biosensor? There are lots of definitions. In 1996, I will just
read it, IUPSC, you know IUPSC International Union for Pure and Applied Chemistry. They
defined a biosensor like this. A biosensor is a self-contained integrated device that
is capable of providing specific quantitative or semi-quantitative analytical information
using a biological recognition element, which is in direct spatial contact with a spatial
transduction element. I will explain it later, but let us see another definition which was
recently published. A sensor that integrates a biological element with a physiochemical
transducer to produce an electronic signal proportional to a single analyte, which is
then conveyed to a detector. And then a very simple one, any device that uses specific
biochemical reactions to detect chemical compounds in biological samples that can be called a
biosensor. So, from here we take home message is that
any device which uses specific biochemical reactions from all the, I mean, this is the
essence of all these definitions and it detects something in biological samples that is all.
So, why it is important actually? Because we are polluting almost everything. I am only
talking about biosensors here and you can see that be it air, water or food. We are
putting lots and lots of pesticides to grow more and more food, but at the same time these
pesticides, the remnants of them are still there in the food. And what will happen if
we eat them? Lots of diseases, cancer, diarrhea, bacterial infections. You can think of a fish
for example, grown in this kind of water. What will you get? This kind of fish which
is already contaminated, yeah. And what will happen if you eat this? You will be subject
to various diseases. For example, this dog has diabetes. So, diabetes is a very common
disease. So, it is very important to detect them and this is only biosensor. I am not
talking about the sensors in defense or in automation and everywhere else.
You can see that the market overview of it that the global market in billion dollars
is increasing at a rate of about 2 billion dollars per annum. So, by 2024, it will be
about 30 billion dollars US dollars. So, there is a huge potential and requirement for biosensors
and that is why this course. Let us see what are the components of a biosensor.
If you have a biosensor, there should be a surface which interacts with these biological
molecules which senses these biological molecules. This is called sensor surface. So, the molecule
of interest is called analyte. You have a molecule which you want to sense in a matrix
of elements. Say, I have put different shapes here to see that there can be lots of molecules
and from there I want to detect only this kind. To detect this particular molecule,
we need some biorecognition element or bioreceptor. What it does is actually that it leads to
a specific attachment of this particular molecule. It does not bind to anything else. It is very
simple. It is like this pen. You have this cap and you have this pen, ok.
Suppose I attach this here on the surface and I have this pen. It is a free molecule.
It comes and binds in here very specifically. What will happen if I bring this one? Say
another molecule, it does not fit in here. That is how this works. And then we have an
agent called anti-fouling agent. What it does is that it blocks any empty area on the surface,
so that nothing else can go and bind in there. It is called to avoid any non-specific binding.
We put this molecule here which fills up all the empty area apart from this BRE, ok.
Now we have a surface which is very specific to this particular molecule. This molecule
goes and binds in there very fine. What happens next? When it binds over here, it will lead
to change in certain properties. Maybe it will lead to change in pH or refractive index
change or maybe change in mass or heat transfer or many things. It can change to current or
something or something. And whatever change took place here, we want it to be readable.
So, there should be some mechanism which gives us the change in a readable output whatever
change occurred due to the binding of this molecule. So, that is the role of something
called transducer. What it does is actually it translates the effect of this binding into
a measurable signal which can be an optical signal or electrical signal or acoustic or
maybe change in dimensions. I told you that the change in length was there in thermometer.
So, there can be this kind of things and once you have this measurable signal, you send
it to the detector, ok. You send it to the detector. So, this is how you make a biosensor.
Now let us discuss one by one these components. So, let us see analytes. Analytes are chemical
or biological or environmental elements that need to be sensed. So, anything you want to
sense is an analyte. It can be a natural hazard like pesticides, environmental pollutants,
what are urine analytes? Say glucose, you can have say cholesterol, you can have ions,
maybe bilirubin, jaundice, all these parameters in a urine or blood whatever you measure,
these are all analytes. Vitamins, now environmental pollutants, you
have say carbon dioxide, methane and all these gases, industrial waste, biochemical waste,
biological waste. From the hospitals you get tons of syringes and this blood soaked cotton
and all these things. You never know, you want to see something in the food, if there
are pesticides or there are preservatives. Ambient conditions, if there is humidity,
there is change in temperature, change in pressure. Gases, as I told you, you want to
measure oxygen, hydrogen, nitrogen maybe in some samples, hydrogen sulfide, methane. There is lot
of use in petroleum sector. Endocrine disruptors, you are eating something, they can be found in
say cosmetics. You can have lipsticks and these powders and all these creams, they all have this
kind of elements which can cause to endocrine disruption. There can be pathogenic bacteria,
in your water. So, you want to detect them or other food related issues. For example,
you have meat, you never know that it is infected with bacteria or viruses. So, you want to detect
them. So, there is a whole range of elements, I mean, it is a very wide field. You can have
analytes say from defense sector to food, to environment, to water, to blood, everywhere,
ok. And they are very important to us. So, these are the molecules.
Then you have receptors or biomolecular recognition elements. You want to have
certain molecules which need to be attached on the sensor surface and you want them to be very
very specific to the analyte molecule. That is called receptor or biomolecular recognition
element. This BRE or receptor is something which needs to be fixed on the surface of the sensor,
so that the analyte comes and interacts with it. Like I gave the example of this pen. So,
the molecule comes and fits in it. So, this is BRE. It has to be very selective and specific.
It is not like if this is coming and something else can also come and fit. No, that is not
acceptable. So, it can be various types. It can be enzyme based. You know what are enzymes?
So, enzymes are molecules which increase or decrease the reaction rates in the body,
metabolic processes. So, something comes and binds on the enzyme, say a ligand and when it
binds it gives us a byproduct which can be used for sensing. Then you have immunosensors,
antibody and antigen based sensors. What it does is actually that you have either
antibody and antigen. One of you it works as a receptor, one of it works as an analyte and you
use it for this. Then you have DNA based sensors which are nucleic acid based. Also,
you have cell based sensors or tissue based sensors. So, these are various kinds of sensors
people use. And then you have transducers. As I told you that it converts one form of
energy to another form of energy. That is the basic definition. But what it does in the sensor
is that basically it transforms this binding phenomenon or interaction phenomenon from
between the analyte and the BRE into a miserable output signal. That is a transducer.
So, when I say optical sensor, that means, the transducer is optical. It is giving us
optical readout. If I say electrochemical sensor, that means, this sensor has a readout which is
due to electrochemical processes. If it is a piezoelectric sensor, that is the transducer
is piezoelectric. So, you name a sensor, the method, it is basically the transducer, ok.
So, for example, I have shown this diagram here that you can see that a transducer can
be electrochemical, piezoelectric, thermometric or optical. Electrochemical can further be
divided in amperometric, conductometric, potentiometric. What does it mean actually?
It means that, say for example, you have amperometric sensor. What it does actually?
It means that in electrochemical sensor you have electrodes, basically three electrodes. And what
you do is that when this interaction is taking place between the receptor and the analyte,
there is a change in current. So, you measure the change in current and that is why it is
called amperometric. When you measure a change in conductance, it is not you measure, actually
it is the parameter which is mostly getting affected. I will come to you why we need a
parameter which is getting most affected due to this, that I will come to you later.
But for now, take my word that you have this kind of interaction and then there is a change
in current between the electrodes. You measure the change in current that is amperometric. You
measure the conductance that is conductometric and if you measure the change in potential drop
that is potentiometric sensor. In piezoelectric, actually the transducer is like if you attach
something on the piezoelectric crystal, what happens actually? There is a change in mass.
The change in mass leads to change in the vibration frequencies. So, you can measure
this small change in vibration frequencies because this piezoelectric means that it will
change the pressure. So, you transform that into the electrical signal and that is how you use it
for transduction. In thermometric ones, you measure the change in heat and that is how you do.
In optical transducers, basically you measure the change in optical signals. So,
there this interaction took place and this leads to some change in optical properties, ok.
So, we know now the basic components of an optical sensor of any sensor. Now,
we want to make a sensor chip. How do we make it? You take a transducer surface,
you put a cross linker. The job of the cross linker is to attach the biorecognition element
on top. You know, sometimes this element can directly get attached to the transducer surface.
Then you do not need to put a cross linker, but many a times this molecule cannot directly get
attached to the surface. You need something say an adhesive for example, kind of I mean
you can use a word vaguely like adhesive. You need an adhesive to put this molecule on the
sensor surface. After you have put it, now it can catch the analyte, but then there are lots
of spaces which are not filled. It is like this. You have a surface, you put molecules here,
here, here, here, here all these places, good. And these places the analyte can go and bind on
here, but you have lots of space here which is empty here, ok, all this space. So, anything can
go and bind on it. This will create a false signal. So, it is very important for us to block
all these sides, so that nothing can go and bind on here, ok. How do we do that? We choose a
molecule or something which goes and binds in all non-specifically on all these things,
on all this surface. You know sometimes you will have this cross linkers which are unbound like
here. So, you have this cross linkers which are unbound here, here, one. So, what will it do?
Any molecule can go and bind here, which is not binding here maybe it will bind here. So,
the job of this agent is to block all the non-specific binding sites. And once we are
done with this, we get a sensor chip. So, you get a sensor chip which only has this
B-ary open for binding, everything else is closed, ok. And when you have an analyte,
it goes and binds on these sides. It does not go and bind anywhere else. If you have another
molecule, it does not go and bind anywhere. That is how you make a very good sensor surface.
Now, when we come to optical sensors. So, optical sensors are generally based on
measurements of changes in any of the following optical properties. It can be absorbance in
chemical reaction, it can either of reflectance and transmittance, it can be change in refractive
index, it can be based on change in phase shift, it can be change in polarization or it can be
based on change in light energy. So, it can either be a fluorescence based sensor or Raman
based sensor. So, if I write the most general equation of the plane wave incident on any medium
and want to see what kind of parameters are there which get changed. So, you can write it like this
exponentially e to the power i omega t minus k dot x and plus initial phase phi. So, suppose
this is a plane wave which is incident on some medium and you want to see what are the optical
properties which can get changed. So, if it is change in amplitude, then it
will lead to sensors based on absorbance or reflectance or transmittance and many times
change in fluorescence intensity or Raman intensity. And, when there is change in k
or omega or change can be in phi and then this z. So, there are basically these parameters
which can change. So, it can be either of this or this or this parameter or this parameter
and then this parameter, which basically can be monitored and the optical sensor can be
based on any of these. So, as I already told you that when it is
change in amplitude, then the sensor is basically absorbance or reflectance or transmittance
based sensor or basically change in fluorescence intensity or change in Raman intensity and
this k is equal to omega by c into n. So, if this n is changing, basically this k is
changing and it gives the direction of propagation. So, suppose this is in x direction. So, basically
let us say that the wave is moving in x direction. So, this refractive index if changes, then
it leads to change in. So, if this kind of plane wave is incident at that medium and
the wave which came out, if it has different k that will give information about n, if there
was no change in omega. Also, if there is change in omega, which is
like you shine with some light and you are measuring say lambda and then there is a change
in lambda. So, suppose this is intensity versus lambda curve and what you see here
is that if there is a change in lambda, then it is change in wavelength. So, basically
it is like fluorescence or Raman based sensors. This initial phi can also change and that
will also get reflected in terms of the absorbance or transmittance and by measuring the change
in phase, you can say that what is the property of the medium.
Another property is the polarization. So, suppose you send a linearly polarized light
through a medium and then after passing through it, its orientation changes. So, by measuring
the change in orientation of the polarization of light, you can say that what is the nature
of this material. So, many times it can be a Levo rotatory or dextrorotatory material
and depending on the changes in the polarization vector, you can say that what is the property
of that particular material. So, an optical sensor can generally be based on any of these
kind of parameters. Why would we need optical biosensors? Because
they are very small, they are flexible, they are very fast, they are safe because there
is no electrical device to interconnect, they have very good biocompatibility, fibers are
glasses, there are maximum glasses most of the times or you use noble metals silver or
gold which is good for health, you wear it all the time. Disadvantage is that the optical
signal may not be strong enough many times. For example, fluorescence, if you have a single
molecule, the fluorescence is very weak. So, we will discuss this as the course progresses.
So, to summarize my talk today, we introduced what are sensors and biosensors and what are
the components of the sensor, how they work and what are their types and then we discussed
the various optical methods which can be probed optical properties and then we use it for
optical sensing. Thank you.