Using a great model: Fisher-Paykel's BABY LIV (Lung Inflator Generator) we show you what happens to a baby's lungs when getting positive pressure and how changes in interfaces can change that pressure. See what happens when the pacifier comes out!!
THANK YOU SO MUCH to Fisher-Paykel for letting us use their BABY LIV. (This was NOT a sponsored post!)- and especially to Kelsey LeBlanc who helps make everything happen!
For the PDF summary of THIS video, links to relevant articles (+ 50 free Multiple Choice Questions/ Answers and explanations!!), AND a bunch of other stuff, subscribe to our Neonatal Nuggets Newsletter here: https://talatalksnicu.substack.com/subscribe
--------------------------------------------------------------------- Please be sure to give us a like, comment about future topics, and subscribe to this channel to stay up to date on all things NICU! Also, click the notification bell after you subscribe to gain your weekly NICU knowledge as soon as it's loaded! --------------------------------------------------------------------- Dr. Tala is a board-certified neonatologist and has worked in busy level III and IV units for the past 15 years. She has won multiple teaching awards throughout her time as a neonatologist. ---------------------------------------------------------------------- References: Luo K, Huang Y, Xiong T, Tang J. High-flow nasal cannula versus continuous positive airway pressure in primary respiratory support for preterm infants: A systematic review and meta-analysis. Front Pediatr. 2022 Nov 21;10:980024. doi: 10.3389/fped.2022.980024. PMID: 36479290; PMCID: PMC9720183.
---------------------------------------------------------------------------- *Disclaimer*: This video is intended for educational purposes only and while we strive to give the most accurate information, errors may occur. Subsequently, this video should not be a replacement for medical advice.
AI Generated·May contain errors
Hello, I'm Dr. Tala and I've been a neonatologist for 16 years now and the video this week is going
to be a quick one. The lovely people at Fisher and Paykel, thank you Kelsey, have lent us their
BabyLiv which stands for lung inflator visualizer. So basically using this model we're going to show
you what actually is happening in the baby's lung when the baby is on CPAP versus when the baby is
on high flow and how little changes in the prongs or with taking out the pacifier can really make a
difference to the pressure that the lungs are actually seeing. Before I go over the model let
me just say a few words about CPAP and high flow nasal cannula. So remember CPAP has been kind of
the gold standard of non-invasive support for like the last couple of decades and it's been
fantastic because we've been able to give babies the support they need without keeping them
intubated. So decreasing their BPD, decreasing their chance of pneumonia. More recently we've
started using high flow nasal cannula in places where really we'd previously been using CPAP.
So basically kind of younger babies, maybe babies that just got extubated and there are some
advantages of the high flow nasal cannula. So for example we can feed babies on high flow nasal
cannula. Generally babies are more comfortable on high flow nasal cannula. They're less likely to
get a pneumothorax on high flow nasal cannula but realize high flow nasal cannula provides less
pressure and less support to the lungs than CPAP. So ultimately we do need to make sure that the
babies get the support they need and the model that we're about to show you really kind of shows
the differences in the amount of pressures and support that the lung is getting. So I hope you
enjoy this. Okay so meet Liv the lung inflator visualizer and Liv is a or baby Liv was based on
the the anatomy of a 28-week-old. So what this company Fisher and Paykel did was they did a lot
of CT scans looking at the airways of babies who are about 28 weeks and then basically reconstructed
the airway. So here you can see that this is the airway going down to the baby's lung. So this
is basically reflecting the baby's lung. Then depending on what pressures and interface and
whatever we're giving to the baby we can look at the pressure that's being generated in the
baby's lung right here. Okay let's start by doing a trial of bubble CPAP. So here this is the water
it's all gonna fall off. So this is the water supply basically that's going to provide the
bubbles for the bubble CPAP and you can see right now when we don't actually have the mask on the
baby it's not bubbling at all. When you put it on it actually starts bubbling. So at this point
because the pressure circuit is complete we actually have bubbling and the baby is receiving
its bubble CPAP. So you can see here right now we have the bubble CPAP set at a water pressure of 6
and you can see right here that we have about between 6 and 7 being generated on this screen.
A water pressure of between 6 and 7. Look how you can see the baby's lungs oscillating a bit
that's because of the bubble CPAP. So now let's watch this baby breathe. So the lungs are going
in and out and you can see that the pressure really doesn't change too much significantly
which also really indicates that there isn't a huge increase in the work of breathing in the baby.
As you all know with CPAP we have a completely closed system so we don't have escape around the
even if we do have the prongs in place obviously this is the mask here and the pressure that the
baby is getting is very much dependent on having that closed system. So for example let's see what
happens if we take the pacifier out of the mouth. You can see that the pressure has already
significantly fallen. The makers of BabyLiv, Fisher & Paykel, do point out though that in a human baby
we also have the tongue and other areas of resistance that aren't going to allow the
pressure to fall all the way to one or whatever it is here. But so this is a pretty exaggerated
response as we're opening up the system but just be aware at the bedside that if the mouth is
gaping open or obviously if the mask is really loose then the CPAP isn't going to work and we
can tell the CPAP isn't working well because immediately you stop seeing the bubbles and you
stop hearing the bubbling. If on bubble CPAP you stop hearing the bubbles by definition the baby
isn't getting the pressure that it needs to be getting. Okay so what we just showed was the baby
receiving CPAP through the mask. So you can see that this mask goes on here and then the rest of
the CPAP interface goes on like this. So the other thing is as you all know sometimes in the unit
we like to alternate the mask with the prongs and we kind of feel that that gives less pressure at
all times to the nares. Generally although we will say that the babies kind of prefer the
mask more than the prongs because when we are putting the prongs in on CPAP it's very important
that the prongs completely obstruct the nares because that's how the pressure is being delivered.
So the outer diameter of the prongs for CPAP is going to be very similar to the outer diameter
of the nares. The other important thing that you need to know when you're using prongs on CPAP
is that we have this interface with the inspiratory limb and the expiratory limb
very close to the nares itself. So the importance of that is that with a completely obstructed
system, and we can put the baby's pacifier back in, with a completely obstructed system obviously
we need a way for the carbon dioxide to come out as well. So here we have the oxygen coming in
and with this expiratory limb we want the CO2 to be able to go out that way since like I said
before there's really no other mechanism with a completely closed system to actually ventilate or
get the CO2 out. Okay another thing that I want to show you on this is that sometimes we use like
a regular nasal candela interface to actually deliver CPAP by hooking it up to the rest of
CPAP machinery. So let me just show you kind of what happens here and there's two big things that
I want to point out. The first one is is that even as we put the prongs in the nose and generally
remember that these prongs are kind of normally made to have a little bit of leak around them and
we'll talk about that more in a second about the importance of that when you're giving high flow
nasal candela. But let's then put on the CPAP machine and again we are giving the same pressure
here so we're giving a CPAP of six here so on that you can see that the CPAP the pressure that's
actually being delivered to the lungs is much lower than that when we're on the prongs. So what if we
really try to occlude those prongs further like that? So yes when we do occlude the prongs kind of
as much as we can then that pressure does go up but then we run into another problem and the problem
is is that we are now using this very skinny tubing before we reach the rest of the CPAP
interface to not only provide the oxygen but to actually exhale the CO2. Normally when we're using
a nasal cannula the CO2 is kind of coming out around the leak as well but here we are now
dependent on the ventilation or getting rid of the CO2 through this little tubing. The expiratory
tubing is still quite a distance away so we do worry when we use this kind of cannula setup
with a CPAP interface that we are not adequately being able to expire the CO2 even if it's such a
tight fit that we are being able to get good CPAP pressures. Right now let's go over when a baby is
on high flow. So this is the high flow nasal cannula you can see that these prongs are much
smaller and these are appropriate size prongs for this baby. So really what a lot of companies say
is that the outer diameter of these prongs should be about 50 percent of the outer diameter of each
snare so much smaller. By definition we do want a leak on these babies because we want a leak part
of that makes it a lot more comfortable for the babies but also it's important to be able to
actually get the CO2 and to ventilate these babies through by having the leak. So we've got
this baby on this high flow nasal cannula and right now we are running this at three liters per
minute. So you can see that we do have a little bit of pressure given so the pressure that we've
got is like 0.7 so about one not completely nothing so just going up on the pressures
oh sorry on the flow so now we're going up to a flow of seven liters per minute so on a flow of
seven liters per minute we're giving slightly lower pressures. So let's see again if you completely
occlude that more then obviously those pressures are going up a lot so that is a very good reminder
for you all that there is a way and you can see those lungs are like going to explode now just
from this high flow so that's just a very good reminder that if you do end up really occluding
the nares with these prongs you are going to deliver very high pressures and it's going to
take away from what we're trying to do which is comfortably give the support without creating
barrow or volume trauma to these babies. So yes some pressure is given but it's also dependent
on exactly how occluded those nares are. So again let's take out the pacifier and that pressure
falls down considerably.