BIOPOTENTIAL AMPLIFIERS(Part 2) || EC365 BIOMEDICAL ENGINEERING FOR KTU || LECT 7
This video lecture contains
1.Bio potential Amplifiers
a. Chopper Amplifier
b.Isolation Amplifier -Transformer isolation, Optical isolation, Capacitive isolation
#ktubiomedical #ktu #biomedical #ktuececlassroom
Hi all, last class we were discussed about biopotential amplifiers, in that we were discussed
instrumentation amplifier and carrier amplifier. Today we are going to discuss
chopping amplifiers and some isolation amplifiers. So what is chopping amplifier? What are its use?
Chopper amplifiers are used to avoid drift problems that may occur in direct coupled amplifiers.
This circuit uses a chopping device. What are the functions of a chopping device?
Chopping device which converts slowly varying DC to an alternating form with amplitude proportional
to input direct current at phase depends on the polarity of the original signal.
So chopper is mainly used to convert DC to AC. So here it converts slowly varying DC to
an alternating form with amplitude its amplitude proportional to input direct current
and phase depend on the polarity of the original signal. The AC voltage applied is then amplified
by an AC amplifier whose output is then rectified back to get amplifier amplified direct current.
This is the working of the chopping circuit. So here the figure shows
a single ended chopper stabilizing circuit. So the aim of the circuit is to avoid DC offset
voltage present in the output signal. For this purpose first convert DC signal into an AC using
a chopper. So here we have to avoid DC offset voltage present in the output signal. So first
of all we have to convert this DC input to an AC signal using a chopper circuit.
So what are the components used in this circuit? So first of all a low pass filter.
What is the function of a low pass filter? It passes only low frequencies and attenuates all
high frequency components. So the low frequency components derived from the input signal by
passing it to low pass filter R2C2 and R2. Here this is the low pass filter. Low frequency component
in the input signal is passed through this circuit. Next is a chopper. The output of
low pass filter is then chopped using a transistor switch with respective carrier signal from the
oscillator. So this is the chopper circuit. So this is a transistor and this transistor
works as a switch. The signal from the low pass filter is passed to this chopping circuit.
This chopper works based on the carrier signal generated by this oscillator. So next is a
demodulator. The original signal is recovered in demodulator which is again applied for
second stage amplification. So here this demodulator will convert this chopped signal
to the original input signal. So after demodulation the signal is transmitted
to second stage amplifier circuit. So again a low pass filter circuit is used
before given to the amplifier low frequencies components again derived using this low pass
filter. Here also a low pass filter circuit like here R3C3 and R3. This forms a low pass filter
circuit. R3C3 and R3 forms a low pass filter circuit. This low pass filter circuit's function
is before given to the amplifier low frequency components again derived using this low pass
filter. Last one second stage amplification A2. At the input of A2, an high pass filter C1R1 is
used to derive high frequency signal. This is to reduce DC offset and drift of second amplifier.
Here we are using an R1C1. This is a high pass filter circuit.
Its main function is to derive the high frequency signal. This is used to reduce
the DC offset. That's why we are using the high pass filter. The high pass filter is to
reduce the DC offset and drift of second stage amplifier. This is the working of chopper amplifier
circuit. Chopping device is used to avoid drift problems occurs in direct coupled amplifier.
So, what are the advantages of chopper amplifier? It is insensitive.
That means it is insensitive to component changes due to aging, temperature change,
power supply variation and other environmental factors. So, it is highly insensitive.
Next is small offset voltage. There is only a very small offset voltage in the case of
chopper amplifier. Offset voltage is the unwanted voltage in a circuit that is
dissipated in a circuit that is offset voltage. So, in the case of a chopper amplifier,
there is only a very small offset voltage. Chopper amplifier is used to amplify small signal DC
of a few micro volts. It can be used to amplify small DC signals of few micro volts. So,
these are the advantages. Insensitive small offset voltage can amplify very small values
of DC signal that is DC signals in micro volt range. So, that's all about chopper amplifier.
Next is isolation amplifier. What are isolation amplifiers? Isolation amplifiers are used in
medical instrumentations to ensure isolation of patient from power supply leakage current.
So, it is used for isolation of a patient from power supply leakage current. That's why we are
using isolation amplifier. They break ohmic continuity of electric signals between input
and output of an amplifier. So, what is the function of an isolation amplifier? Isolation
amplifier is used to ensure isolation of a patient from power supply leakage current.
So, what is the function of an isolation amplifier? Isolation amplifier is used to ensure isolation
of a patient from power supply leakage current. That's why we are using isolation amplifier.
They break the ohmic continuity of electrical signals between input and output of the amplifier.
Isolation amplifier is used to ensure isolation of a patient from power supply leakage current.
Mainly, there are three methods used in isolation amplifiers.
Three types of isolation amplifiers are transformer isolation, optical isolation,
and capacitive isolation. So, first we have a look on transformer isolation.
It uses either frequency modulated or a pulse width modulated carrier signal with a small
bandwidth of up to 30kHz to carry the signal. It uses either frequency modulated or a pulse
modulated carrier signal with a small bandwidth of up to 30kHz to carry the signal.
It uses an internal DC-to-DC converter comprising of an internal DC-to-DC converter
of 20kHz oscillator, transformer, rectifier, and filter to supply isolated power.
Internal transformer isolation uses either frequency modulated or a pulse modulated carrier
signal with a small bandwidth of up to 30kHz to carry the signal.
Mainly, it uses an internal DC-to-DC converter comprising of an internal DC-to-DC converter
of 20kHz oscillator, transformer, rectifier, and filter to supply isolated power.
Optical isolation uses either frequency modulated or a pulse width modulated carrier signal
with a small bandwidth of up to 30kHz to carry the signal. It uses an internal DC-to-DC converter
comprising of an internal DC-to-DC converter of 20kHz oscillator, transformer, rectifier,
and filter to supply isolated power.
Optical isolation uses either frequency modulated or a pulse width modulated carrier signal
achieved by optical means in which the patient is electrically connected with neither the hospital
line nor the ground line. A separate battery-operated circuit supplies power to the patient
and the signal of interest is converted into light by LED.
This light falls on a phototransistor.
To send the signal across differential capacity barrier.
Separate power supply is needed on both sides of the barrier.
Also signals with a bandwidth up to 70kHz can be conveniently handled in this arrangement.