Color accuracy begins with calibration. In this video, Presto Stantest, Asia’s No.1 testing equipment manufacturer, explains the step-by-step calibration process for the TP 800 Spectrophotometer. Ensure your readings for whiteness, yellowness, and color variation are reliable, repeatable, and audit-ready with proper calibration.
For more information Visit our website:-https://www.prestogroup.com/articles/spectrophotometer-what-it-is-and-how-to-use-one/
Hello everyone.
In this video, we will discuss about how to perform the calibration of spectrophotometer
TP800.
So for performing the calibration, I have this kind of calibrator.
In this, we perform a black calibration and white calibration.
And by these two calibrator, we ensure that our testing result when we perform the color
comparison, it is accurate.
So before performing the test, it is advised that we calibrate our sensor.
So how do we perform the calibration, let us understand.
So this is our calibrator.
So the first part you can see, this is our white calibrator.
The material used here, it is a PTFE.
It is a plastic kind of component which is standardized to perform white calibration
and up to 100% reflectance, it used to calibrate the sensor.
So this is for white calibration.
And for 0%, we used to calibrate from 0% to 100%.
For 0% calibration, we used to use a black calibrator.
So this is the black calibrator.
Here we have to make sure that it is always closed, covered whenever we are not using
it.
Otherwise, if any kind of dirt component will attach on this, then automatically the result
you will see will be varied and your testing result will not be appropriate.
So this we have to make sure.
Now I will show you that how to perform the calibration.
Now we will switch on the equipment and we will understand that how to perform the calibration.
So from here, I will switch it on.
Here is the measuring key.
Now I will open the cover of this sensor.
This is silicon photodiode.
Now how to go in the calibration mode?
We have to go in this menu.
We will go in calibrate.
We can see we have many options.
Right now the purpose of this video is to understand the calibration.
We will go in calibrate.
Now I want to perform the white calibration.
So I will go in white calibration and I will open this, the white calibrator.
And this sensor I have to place over the calibrator.
And then either I can press this OK button or the measuring key.
So I will press this OK.
You can see white calibration success.
If I will keep it in like this in air or if I will move it here and there, keep it loose,
then the calibration will fail.
So it is very important that the sensor is properly centered to this white calibrator.
So this is completed.
Now we will do the black calibration.
I will cover this.
And now we will go in black calibration and we will go in OK.
You can see black calibration success.
So this is how we perform the calibration of the TP800.
So by this you can make sure that your results are accurate.
Thank you so much.