How to Perform Calibration of Edge Crush Tester | Step-by-Step by Presto
Prestogroup Online
1 follower
2 Video Views·Apr 29, 2025
Accurate results start with accurate calibration. In this video, Presto Stantest, Asia’s No.1 testing equipment manufacturer, shows you how to perform the calibration of an Edge Crush Tester (ECT). Learn step-by-step how to ensure your machine gives consistent, precise results—critical for corrugated box testing and packaging quality control.
For more information visit our website:- https://www.prestogroup.com/products/edge-crush-tester/
Hello everyone.
In this video, we will understand that how to perform the calibration of the load indicator
in the edge crush testing machine.
So we will use a dead weight in this.
So when we perform the calibration, we have made one more video on tensile tester load
cell calibration.
So the procedure of the calibration is dependent on the how the manufacturer has made the machine
and the internal settings.
However, the principle of the calibration is same throughout the area and to the industries.
So we will understand that principle only here, okay.
So the principle is that we have to first zero the indicator, we have to tear the whole
reading, then we have to apply dead weight and then we have to feed the dead weight value.
If it's a 2 kg, 5 kg, 10 kg as per the load cell and then we have to just take out the
load cell, basically the dead weight from the compression plate and then we have to
just see that whatever the dead weight is, if it's a 5 kg dead weight in the indicator
also it is showing same or not.
So this is how we perform the calibration.
Now we will perform the calibration.
So we have to go into the internal setting, we will set the password first, we do not
have to touch this.
So this is the area where we have to make the changes, okay, when we will apply the
load.
So we have to make sure that when we are here on this cal mode, before this only we do not
have to apply the dead weight.
When we are coming into this section, now we have come into this calibration mode, now
we have to apply the dead weight, whatever dead weight we are using.
So now I will be applying the dead weight.
So before that what we have to do, I will take out this upper gripper, the upper compression
plate and I will apply the dead weight.
So this is a 5 kg dead weight with which I am calibrating the load cell.
So now we have to feed the dead weight which we are using.
So it's a 5, so already it's 5, but let me just show you that how we have to do.
This is how we have to go into this.
And then we will go to next option, this is factor, we do not have to change this, no
need to change the dot point and then again we will move and this is the main indicator.
Now let me show you that our machine is calibrated, load cell parameter.
So I have just taken out the dead weight, now again I am putting the dead weight.
So you can see the indicator is giving the appropriate reading.
So this is how we calibrate the load cell in the compression mode.
Right now the load cell is here.
Thank you so much.