Lecture 18 : AFM and Recap Raman and IR Spectroscopy

IIT Kanpur July 201
20 followers
Aug 9, 2024
[Music]
so welcome back to the lecture
18th where we will be talking about a
little bit more about the AFM and we'll
have a recap of Raman and a
spectroscopy because I realized that
um we'll be talking a lot while talking
about most of these
bioelectrochemical and Optical sensors
about the interaction of light and
matter
so I decided that we will
have a reasonable recap so you can
appreciate once again the power of some
of these extraordinary spectroscopic
tool because
now as I have been mentioning for last
three or four lectures that all the
instrumentations which are happening
they are all
multimodal operations you have Optical
you have
electrochemical you have a
spectroscopic you
have some of the electronic features of
molecules you have the mechanical
aspects of it in in one platform
so let's recap we finished the last
class with the tapping AFM where we
showed
that photothermal expansion due to
sample absorption causes an increase in
the amplitude as you could see and that
is what is being recorded and that's
what helps in espcially understanding
the biological membrane sensitivity and
chemical Imaging now from here we move
on to the contact resonance now here we
will talk about say for example here is
a surface and here you have the beam
falling and the beam is reflecting right
now as soon as this C tip moves so there
is a change in the frequency as you
could see here there's a change in the
frequency so this is called Contact
resonance AFM or CRF AFM is a dynamic
contact technique so again you Rec
and I talked to you about
contact non- contct
and
Inter mittent modes so you always have
to remember which mode are we talking
about in this case we are talking about
the contact mode that makes use of the
vibrational behavior of the C liver
while the tip is in permanent contact
with the sample generally an increase in
the the sample stiffness prompts an
increase in the contact resonance
frequency it is almost like you see for
example this is the harder surface right
you hit something on this you'll get a
much bigger Force versus you hit a
softer surface say for example suppose
you jump on a
sand you won't get hurt but if you jump
on a concrete floor you will get hurt
because it will force back at a higher
Force right so the reverse force will be
higher the counter force will be higher
exactly the same way when I am
mentioning out here that generally an
increase in the sample stiffness prompts
an increase in the contact resonance
frequency that means if this particular
part is higher than
the contact frequency will become higher
okay so this is very important for you
to understand and that's how you can
figure out whether the surface
is softer or a surface is
harder so especially when we talk about
soft and harder surface mostly these
kind of surfaces
are biological
materials and much of the harder surface
are inorganic
systems so depending on what kind of
samples you are looking into you have to
make a j judgement and based on that you
can develop the
topography CR
frequency contact resonance frequency
and elastic modulus of the particular
sample and if you look at CFM the local
elasticity of the surface is prob we
observing the shift in one of the igen
mode resonance frequencies of the FM
prob brought into contact with the
surface the resonance frequencies of the
amm C changes in accordance with the
elastic stiffness of the probe sample
contact so if you look at it here it
is right so depending on the frequency
changes so here you could see the change
which is happening out here right
schematic of the CFM measurement setup
with a mechanical modulation from the
base of the canli right what you see as
the crfm frequency spectrum highlighting
the shift of resonance frequency in the
first to igen mode from out of contact
to contact on Silicon surface SI is here
is the Silicon surface so these
are the two igen mode which are being
observed out there okay so and as you
could see in the air and this is hitting
the surface so it is basically moving
and hitting the surface and moving
hitting the surface
like now when we talk about the CRM
based on the can Dynamics the measured
CR frequencies are converted into
contact stiffness so you could figure
out what is the stiffness of the surface
which are further used to determine the
elastic modulus of the sample tested by
applying an appropriate contact
mechanics dision to the two sample
contact this poses a few challenges in
terms of unavailable geometrical
parameter the tip shape contact radius
different surface mechanics across
material addition Vis elasticity and the
need for analytical contact mechanics
model for various sample geometries the
analysis of the contact
mechanics is done either by an
analytical model or numerical
calculation in the absence of an
analytical solution also the accuracy of
CFM measurement is greatly enhanced by
using reference material with successive
measurements on the test material and
reference material of known elastic
moduli this procedure can be circumvent
some parameters that are not available
for measurement so basically what of the
easier way is that you use a system
about which you have a fairly good idea
known system so the known system becomes
your reference frame once you are using
the known system as it mentioned also
using the reference materials out here
with a successive measurements you will
be able to detect a newer material in
the successive experimental platform
right from here today we'll move to the
part two of this lecture which is the
recap of the Raman and the IR
spectroscopy because these are the two
systems which are very widely coupled
with AFM
systems for
multimodal
Imaging so let's recap of what is Raman
and what is IR
spectroscopy before I start what is
Raman and what is IRS
spectroscopy try to understand few
simple
things in terms of interaction with
matter for example this is matter right
and these are the atoms which are
constituting the matter if a light
falls either some of the light will bend
and pass through like this some of it
will get
reflected some of it will get
scattered some of it will get absorbed
so either we will
observe scattering
or
reflecting or bending and
refracting
or absorption or they will create
uh vibration in them they will led to
vibration of
atoms and this vibration of atoms is the
basis of both these spectroscopic
techniques which falls under IR infrared
and Raman spectroscopy both are
vibrational spectroscopy it is a
vibration of the atoms in a matter while
you are trying
to capture the unique vibrational mode
and
assign it a chemical nature this is that
interaction
of light and
matter and the rise of one of the most
beautiful field of analytical chemistry
the spectroscopy okay so having said
this let's move
to what is Raman and what is IR
spectroscopy when you talk about Raman
we always have a pride that this was all
done by CV Raman in the year 1928
so Raman spectroscopy is a popular
technique for the analysis of molecular
structure and is considered
complimentary to infrared spectroscopy
or IR spectroscopy Raman spectroscopy is
based on Raman effect which was first
identified by Indian physicist CV Raman
in 1928 the Raman effect is based on
scattering of Light which includes both
elastic rally scattering at the same
wavelength and as the incident light so
there are two three things which could
happen so this is the incident light so
it is scattered at the same wavelength
as v0 as v0 which is called the rale
scattering okay as the incident light
and inelastic which is the Raman
scattering at different wavelength due
to molecular
vibration so these are Raman scattering
which are not at v0 V 0 is the initial a
so they are going out at a different uh
wavelengths so these are the two
different wavelength anti- Stokes and
Stokes do not need to get into the
details of that but just understand
Raman scattering is about million time
less intense than rale scattering
therefore to obtain Raman Spectra it is
necessary to prevent this part is very
important that is very very necessary to
prevent the rale scattering from
overpowering the WEA Raman catc so
unless you can deduct this you will not
be able to detect this so easily because
this is pretty pronounced R scattering
so Raman Spectra are measured by
exciting a sample using a high intensity
laser b with resulting scattered light
being passed through the
spectrometer the Raman shift is energy
difference between the incident light
and the scattered light in
the the resulting Spectrum the vertical
axis is is uh the intensity of the
scattered light and the horizontal axis
is the wave number of the Raman shift
which is shown in cm to the^ minus one
and if you look at the Raman shift what
Raman shift is all about this is a
classic example of sulfur so here you
have the rally scattering so you have to
as I was
mentioned this is something you have to
be
careful because this sees the maximum
what comes out of it and this is the
anti- Stokes and sto ran shift Raman
shift is associated with the two
different energy band the shift at
wavelength higher than that of the
incident light is termed as a sto
scattering the shift at the wavelength
lower than that of the incident light is
term as anti- sto cattering so if you
remember V 0 plus if you look at it V 0
+ V and V 0 minus V so this is Raman
scattering where there is an addition
and there's subtraction with respect to
v0 so there is an increase and there is
a decrease and these are the two things
what you get in the Raman Spectra so you
will always see the numbers on one side
will be negative the other side will be
positive as you could see here ran shift
positive and on the other side will be
corresponding negative okay once again
the Raman shift is associated with two
different energy bands the shift in the
wavelength is higher than that of the
incident light is termed as Stoke
scattering the shift at wavelength lower
than that at the incident light is
termed as anti- sto scattering and as an
example Raman spectrum of sulfur
measured with an excitation wavelength
of 532 nanometer green laser is shown
figure Stoke scattering is observed in
the lower wave number longer wave number
region and anti- sto is scattering in
the higher wave number shorter
wavelength region typically higher
intensity sto is scattering Peaks are
used for analysis but anti- sto Peaks
are also used so remember generally when
we report we report as the higher
intensity ones but sometime the lower
one are also used this is how the Raman
shift happens so v0 plus and V 0 minus
which is Stoke and anti-
sto now what is the difference between
Raman Spectra and IR spectroscopy this
is another very interesting part because
many of times I still remember I always
used to get like I used to almost think
think at one point that they are all
almost same they are all vibration but
it's not really true there are very
distinct differences and today we are
going to nail down all those distinct
differences right so if you look at it
both Raman spectroscopy and a
spectroscopy are based on molecular
vibration as I mentioned these are both
are based molecular lustrated below out
here what is being lustrated out here
right infid spectroscopy is based on
absorption of light energy this is very
important for you to understand
corresonding to the vibrational energy
of the
molecule where it vibrates it actually
absorbs the light right Raman
spectroscopy is based on scattering of
incident light at an energy shifted by
vibrational energy H new of the molecule
so this is the very basic
difference one
is suppose there is say for example
assume the atom as a ball out here one
light is falling on it it get absorbed
and it starts to vibrate the other one
it get scattered and it is vibrate it is
almost like you hit the
ball and ball is vibrating or you grab
the ball you're absorbing the energy and
still it is vibrating so these are the
very very fundamental difference so if
you look at the look at the slide you
will realize this
is IR where it is getting absorbed out
here and this is the Raman which we have
already talked about and Stokes and
anti-stokes right so this fundamental
difference has to be very clear for all
of you that these two are different now
when you talk about ran versus a further
though both are forms of vibrational
spectroscopy ir and differ in some
fundamental aspect as shown in the next
figure IR
spectroscopy is based on the fact that
molecular absorption at specific
vibrational frequencies causes change in
dipole
moment this is the second key difference
High spectroscopy is based on the fact
that molecular absorption at specific
vibrational frequency causes change in
dipole moment so this change in dipole
moment could be used for detection Raman
spectroscopy relies on the change in the
polarizability of the molecule at a
frequency Raman shift at which the
molecule scatters
radiation IR Spectros scopy is sensitive
to heteronuclear functional groups
vibrations and polar bonds especially
hydroxy and stretching in water Raman
spectroscopy is sensitive to homonuclear
molecular Bond such as carbon carbon
carbon carbon double bond carbon carbon
triple bond these are good for all the
Raman spectroscopy or if you go for
heteronuclear then you should go for the
IR mostly mostly preferentially this is
how they look like and if you look at
the chart this will give you a overall
idea about the upper one what you see
out here this is all the IR absorption
at different Zone you could see so as I
mentioned these are all heteronuclear
hydroxy nitrogen and
hydrogen ch3 ch2 CH and
then
phenolics carbon nitrogen bonds so you
look at
different part you could see the
different kind of functional groups
which could be detected whereas in
drumman you see mostly mostly they have
a slightly different kind of orientation
in them and this is something which is
always used when you are getting the
data from IR or the Raman you have to
really refer to the chart that where
possibly be and this has been developed
through Decades
of trial and error and identification
and understanding the
exact molecular reason why they show
that kind of absorption at that
particular zone or why they show that
kind of polarizability at that zone why
at that zone they show dipole moment in
the case of um IR versus why they show a
polarizability at that particular
situation of Raman Spectra okay so from
here if you look at the complimentary
relationship of Raman and a there is a
very interesting thing so the upper one
what you see is the a Spectra in red and
you have the Raman Spectra in blue they
almost look similar but they are not
similar there are fundamental
differences a comparison of a Spectra
and Raman Spectra of amino acid line is
shown in figure the intensity of the two
Spectra exhibit mirror symmetry so a and
Raman Spectra often considered
complimentary but they are different in
the types of physical phenomena they can
measure
in Iron measurement the spectral
intensity depends on the size of the
dipole moment as I mentioned they have a
dipole moment for vibrational modes for
bonds such as carbon to oxygen and
oxygen to hydrogen on the other hand
ruman spectroscopy the intensity depend
on the degree of
polarizability so this part is very
important degree
of as I was mentioning in the previous
slide
which is essentially electron volume for
vibrational mode and bond like sulfur
sulfur carbon carbon carbon nitrogen you
could get a very characteristic
signature for them so you're realizing
that why it is extremely essential that
we understand these small molecular
signature in order to venture into the
world of biod
detection because these are small cues
which will help us to detect what kind
of things what we are trying to discuss
okay so I will close in here and then
next class we'll move and we'll talk a
little bit more about raman's microscopy
thank you
[Music]
[Music]
[Music]