KTU BIOMEDICAL ENGINEERING || EC365 || MODULE 2 || LECT 11 || BLOOD PRESSURE
This video lecture contains
1. INDIRECT METHODS-Palpation method, Oscillometric blood pressure method, ultrasonic pressure measurement
2. Direct methods - Percutaneous insertion, Catheterization,Implantation of transducer
For the full video of this course : https://www.youtube.com/playlist?list=PL2UV2EJdMQmh3GlNha5gwN8IsIyjOKcGh
#ktubiomedical #biomedicalengineering #ktu
Hi all, today we are going to learn about indirect method and its continuation.
In the last class, we were discussing about blood pressure measurement.
We have learnt some things about blood pressure measurement.
Today, we are going to discuss about indirect method and its continuation.
In the last class, we were discussing about
pressure measurement and its direct and indirect classification.
In the last class, we have learnt about osculatory method.
Today, we are going to complete the remaining portions of the lecture.
Let's start.
In indirect method, the next one is the palpation method.
Palpation method is to touch the patient's body parts and diagnose.
How is pressure sensing done via palpation method?
Here also, like before, we connect the sphygmomanometer to our hand like this.
We measure whether it was a stethoscope or not.
Here, it is not like that.
We place the pressure cuff on the upper arm.
After placing it, the doctors hold it in their hand like this.
Where? In the wrist.
In the wrist, there is a radial artery.
We hold it like this in the radial artery.
After holding it in the radial artery, the pressure pump is used to increase the pressure.
The pressure is inflated in the cuff.
When the patient holds the cuff in the radial artery, the blood flow is correct.
The pulse is detected correctly.
The pump is used to increase the pressure.
At a particular point, the patient is unable to get the pulse.
The point where the patient is unable to get the pulse is called systolic blood pressure.
Instead of using a stethoscope, the doctor uses the patient's hand to check the pulse in the radial artery.
This is the palpation method.
When the patient holds the cuff in the radial artery and increases the pressure, the patient is unable to get the pulse.
At a particular point, the patient is unable to get the pulse.
This is called systolic blood pressure.
The disadvantage of this method is that only the systolic pressure can be detected.
The diastolic pressure can be detected by palpating.
This is the main disadvantage of this method.
There is an indirect method.
We have covered the palpation method.
In the indirect method, there is an osmometric blood pressure measurement.
The osmometric blood pressure measurement is similar to the sphygmomanometer method.
There is a slight variation in the pressure.
When we inflate the cuff, the pressure is applied to the artery.
There is a slight vibration in the artery.
The vibration at this point is unstable.
This is the process that correlates the pressure variation with systolic pressure.
We can see this figure in the right.
There is no oscillation at a particular value when the pressure is applied.
There is no oscillation at a particular value when the pressure increases.
The point where the oscillations are not there is called systolic pressure.
When we bring the applied value little by little,
you can see an increased value.
The increased value is called mean atrial pressure.
Mean atrial pressure is the peak value of the oscillations.
Atrial pressure is the value when the value is decreasing.
Diastolic pressure is the value when monitoring is required.
There is a microprocessor in this.
The microprocessor periodically inflates and deflates the cuff.
This is how blood pressure is measured in oscillometry.
Next is ultrasonic pressure measurements.
Ultrasonic waves are used in this.
Ultrasonic waves are acoustic waves.
The sound range of acoustic waves is higher than human hearing.
Ultrasonic waves are used in this.
Doppler shift is used in this.
Acoustic waves or ultrasonic waves fall on a moving object.
It reflects from the moving object.
It reflects due to a variation in frequency.
This is the property we use in ultrasonic pressure measurements.
We have placed piezoelectric ultrasound sensors in this.
The sensors are placed on top of the artery.
This piezoelectric ultrasound sensor senses Doppler detection.
This produces a variation in frequency corresponding to the blood flow.
This produces a Doppler shift in frequency called delta f.
This produces a frequency content called f plus delta f or f minus delta f.
This is the corresponding corot-cuff sound to the blood flow.
This is how ultrasonic pressure measurements work.
These are the indirect methods.
Now let us see the direct methods.
There are three methods.
Percutaneous insertion, catheterization, implantation of a transducer in the vessel or in the heart.
This is a direct monitoring method.
This is a highly accurate method.
Percutaneous insertion, catheterization, implantation of a transducer in the vessel or in the heart.
Let us first look at percutaneous insertion.
Percutaneous insertion is the process of injecting a transducer into a tissue.
Local anesthesia is given for percutaneous insertion.
A hollow needle is inserted into the vessel.
A catheter is inserted into the needle.
A metal guide is used to insert the catheter.
A catheter is inserted into the vessel.
A metal guide is used to insert the catheter.
Blood pressure is directly sensed by the transducer attached to the catheter.
A catheter is inserted into the vessel.
The next method is catheterization.
A catheter is a long tube that is introduced into the heart or major vessel by a way of superficial vein or artery.
A catheter is a tube that is introduced into the heart or major vessel by a way of superficial vein or artery.
Blood pressure can be measured in two ways.
The first method is by inserting a sterile saline solution into the catheter.
The second method is by inserting a sterile saline solution into the catheter.
A transducer amplifies the electrical signal of the inserted pressure and displays or records it.
A device is placed at the tip of the catheter.
A transducer is placed at the tip of the catheter.
A transducer amplifies the electrical signal of the inserted pressure and displays or records it.
A transducer is placed at the tip of the catheter.
A transducer amplifies the electrical signal of the inserted pressure and displays or records it.
A transducer is placed at the tip of the catheter.
A catheter is placed at the brachial vein.
A catheter is placed at the femoral vein and at the femoral artery.
Implementation of a transducer in a vessel or heart.
Thank you for watching.
We will see you on next lecture.