Learning Curves Spirometry Video: 6: Valid Spirometry Tests – Acceptable and Repeatable Maneuvers

CDC
258 followers
Apr 28, 2026
This video was originally released in July 2017. Learn what to look for and how to properly interpret a valid spirometry test . This is part of a 7 part series on how to correctly administer a spirometry test using standard professional practice guidelines. NIOSH (DHHS) Pub. No. 2017-167.
In the last module you saw how to perform spirometry and studied normal
curve shapes. Before we study common test errors, let's watch as our instructor
teaches a trainee about what constitutes a valid spirometry test. So on the
flow volume curve on this on the left side, we're looking for a sharp peak just
like this poster has. A sharp peak and it comes right straight back down to
baseline. It has about a 45 degree angle and this is generally a normal test. It has
that 45 degree angle and this is flow. It only tells us flow, the amount of effort
that the subject put out to blast the air out. And then across the bottom it
tells us the volume, how much came out. And we can see six liters there. And then
on this graph you can see time. It starts at zero and goes all the way over to 16
seconds. In the first second they blast the air out and then come up and
make a curve and it shows us the end of test very well. The first one shows the
absolute start of test and the other, the volume time curve shows a really good
picture at the end of test. Okay, we have another test to look at. This is a
smaller volume. Good test and it's continuing. The screen blew up for us
there. And you can see this person has much less air than the one in the
previous test where there were six liters. But it's still an acceptable
test because there's a sharp peak on the peak flow. It comes down all the way to
the baseline. It has a little bit of scooping and that indicates there may be
some disease process going on. But it's very pointy on top which is a good
indicator of effort. And this person takes a little bit longer to get their
air out. They went all the way to 10 seconds but they still got a plateau. So
even though the volume is smaller it's still an acceptable test because they
met all the acceptable criteria. There's no cough, there's no hesitation, there's
no extra breath. None of the errors that you'll see on this poster are on this
curve. But it's so it's still a good test. Now you need to learn how to check
each of the curves for errors. Some of these errors can be seen while the
subject is doing a maneuver. Other errors are only seen after the maneuver has
been performed. Errors may or may not trigger an error message on the
spirometer screen. Technicians must be trained to recognize and correct errors.
Zero flow errors can occur in flow type spirometers. Some flow spirometers set a
zero flow baseline before each test. While other machines set zero flow
before each maneuver. These errors can occur when the system is setting the zero
flow baseline or when the spirometer handle is being moved around or exposed
to airflow before or during testing maneuvers. Don't wave it around or make
any gestures with it because it's highly sensitive. Okay just kind of keep it
right up there close to your cheek. There you go. Ventilation systems or fans may
cause air currents that are not obvious but may contribute to a zero flow error.
All of these maneuvers must be rejected. Zero flow errors can be either positive
or negative. Positive zero flow errors cause over reporting of expiratory
volumes. The volume time curve continues to climb never reaching a plateau or
flattening even after expiration stops. Either way the maneuver must be rejected
when zero flow errors occur as it alters the volumes being reported. The corrective
action for flow errors is to re-zero your spirometer. Okay now just take your
opposite hand take the palm and block the back of the tube for me here for one
second. There you go and we're going to calibrate the machine. Negative zero flow
errors cause under reporting of expiratory measurements. The volume time
curve descends downward. Coughs may occur at any time during the exhalation but a
cough in the first one second may alter the FEV1. Reject all maneuvers with cough
in the first second and document this in the health record. Cough at the end of
exhalation can be a problem if the worker stops exhaling prematurely before
emptying as much air as possible from the chest. This alters the FVC. These
messages mean that the extrapolated volume exceeds 0.15 liters or 5% of the
FVC whichever is larger. A maneuver exceeding these extrapolated volumes is
unacceptable and should be rejected. A poor initial effort or submaximal blast
occurs at the start of a maneuver. A poor or submaximal effort is most clearly
seen as a small rounded peak or no peak on the flow volume curve. Coach the
worker to blast out harder. In the next one see if you can blast harder real
hard at the very beginning. Are you ready? Hang on a minute. Put the mouthpiece
right by your mouth. Big, big breath in. Put it in your mouth quickly. Blast! Now keep
going, keep going, keep going, keep going, keep going. You're doing great. That one
was much better. Variable effort occurs if the worker does not exhale with
maximal effort throughout the maneuver. There is a reduction in peak flow on the
flow volume curve which will be followed by dips or wobbles. In addition
the volume time curve may tilt to the right. Coach the worker to keep blowing
out with maximum effort until told to stop. A partially blocked mouthpiece
causes the speed of exhaled air to be reduced. This can be caused by several
factors. Wobbling curves may appear on both the volume time and flow volume
tracings. The flow volume curve may have a reduced peak with portions of the
graph flattened out. Listen for any noisy exhalations that may indicate air
escaping between the mouthpiece and the lips. Instruct the worker to place the
mouthpiece between the teeth and on top of the tongue. Have the worker demonstrate
for you. Securing dentures is preferable to removing them. Lightly biting on but
not collapsing the mouthpiece may also help. If the dentures are poorly fitted
remove them during testing. Early termination or no plateau occurs when
the worker does not exhale completely. The volume time curve continues to rise
and does not plateau. If the maneuver is terminated early enough you may see a
sharp drop to zero flow on the flow volume curve. The worker should be
coached to keep blowing until instructed to stop. Leaks can occur at any time. An
alert technician can usually hear leaks if there is not a tight seal around the
mouthpiece. In a flow type spirometer a leak causes the volumes to be lower but
can't be seen on either the volume time or flow volume curve. In a volume
spirometer when a leak is large enough to be visualized the volume time curve
will descend after a short plateau is reached. The flow volume curve will move
backwards towards the zero volume at the end of the maneuver. Small leaks
cannot be visualized but may be detected during the volume accuracy check. Glottic
closure or breath hold causes air to stop flowing before the lungs are
completely empty. The volume time curve will demonstrate an abrupt horizontal
line indicating that cessation of air flow has occurred. The flow volume curve
will drop sharply down to the zero flow. This can occur at any time and is more
difficult to identify later in the maneuver. Some people might call this a
glottis closure but that's kind of rare. It does look very similar but generally
it's on every curve in a test. Every curve will have the same effect at the
same spot if it's true glottis closure. However if they're holding their breath
or they've stuck their tongue in the mouthpiece then you get the same effect
as this one. So it could be glottis closure but not likely because it's
only on one trial but it's almost surely either holding their breath or they
put their tongue in the mouthpiece. Extra breaths are usually taken near or at the
end of exhalation when a worker is air hungry. Multiple inhalations can be taken
in through the nose or around the mouthpiece. One or more extra breaths are
easily identified as small steps added onto the volume time curve. The highest
measured FVC is inaccurate when this error occurs. A maneuver with an extra
breath must be rejected. This one has many many extra breaths. This person
appears to me they did maybe one curve that was good and the technician told
them to keep blowing and they were trying to comply with that but didn't
have nose clips on so they had a special skill. They were able to take air in
through their nose as they were blowing and that elevated the FVC to a hundred
and sixty percent of predicted eight liters and eight point three three
liters. There's nothing valid about any of these tests but there's no machine
that I know of that will signal that this is an invalid test. A maneuver is
considered acceptable when the worker performs all aspects of the forced
expiratory maneuver without errors.
Okay Tyler, whenever you're ready, chin up, big deep breath, blast it out, blast it, good job, good job, alright, that's really nice, and stop.
After each maneuver, check to determine whether it is acceptable before asking
the worker to perform another maneuver. If errors are found, take corrective
action before proceeding. A minimum of three acceptable maneuvers are required
while a maximum of eight efforts can be attempted to produce a valid test. Now
what we have here, we have four very nice maneuvers, you did fantastic. We have two
that are very close to each other, okay, and then we have another two that are
very similar as well. Your largest one was actually your second trial, but you
do have to, the doctors will be able to take a look at this information and
they'll be able to give you good results because you gave us a repeatable test.
Okay. So you did fantastic. Thank you. Once three technically acceptable maneuvers
free from errors are obtained, repeatability or consistency of the
results for FVC and FEV1 must be assessed. Determine repeatability by
calculating the difference between the highest and the second highest values
for both FEV1 and FVC. The difference in values between the largest and the
second largest values of FVC and FEV1 must be less than or equal to 150
milliliters or 0.15 liters to satisfy test repeatability. The largest and the
second largest values must be obtained only from curves that are acceptable.
When the worker has produced three error-free maximal efforts and both the
FVC and the FEV1 are repeatable, the test may be concluded. Failure to achieve
repeatability is most often caused by submaximal inhalations, but it may
sometimes occur due to ventilatory impairment, since it may be harder for
sick people to repeat values at maximal effort. The goal of the test session is
to obtain accurate repeatable results. However, repeatability may not be
possible to achieve in some cases. Okay, let's go ahead and try it again. Adam, nice
big deep breath, blast it through. Okay, Adam, now just try to relax and stay tall
for me, Adam. There you go. Almost there. Looking good. Keep going. Keep going.
Almost there. Keep going. And stop. Good job. If the subject is unable to repeat
results, it is important to document this information for the test interpreter. The
next module reviews calibration check reports and test reports in detail.
Technicians need to be familiar with the important elements that should be
included in reports.