Welcome back students. So, as we mentioned in the previous video, like video lecture
that in this particular lecture, we will be looking at the equipment itself, the equipment
that is used for tensile testing, which is UTM, Universal Tensile Testing Machine.
So, let us start with what are the important components of a testing machine. So, let me
draw a simple schematic of how a UTM looks like. So, there is a frame, it is supposed
to be vertical, although my drawing may not show it exactly like that. So, this is a frame
and this is called a double vertical frame. There can also be a single vertical rod for
supporting it and it all basically determines what will be the total cost of the equipment.
And in here, you would have a testing jaw, somewhere over here, this is the bottom jaw,
the top jaw and this top jaw would be connected to a moving part, which will be called a crosshead.
So, this part that we are looking at, this is the frame, this is one important component,
the actually it should be called load frame. And you would see that the quality of data
depends a lot on the quality of the frame that you are using. Then the other important component
is the crosshead. So, you see this is the moving part, this one usually happens to be stationary,
but this also acts like a crosshead and this is also a crosshead. So, crosshead is the one that
moves up and down to pull the sample and in which you would place your load cell.
So, before I come to load cell, let me write down crosshead. So, this is another important component.
Now, you are pulling the sample, but you need some device which would actually measure what
is the total load acting on it and that device is nothing but a load cell. So, usually,
you would have load cell attached to this, which would measure how much load is acting.
And as a preliminary data, the amount of movement the crosshead makes from there, you
estimate the elongation of the sample. So, these are the 3 most important part or the
units of the loading structure. So, these are the 3 loading units.
Apart from that, you will also have testing grips. So, depending on whether you have a flat sample,
a round sample, a threaded sample, you will have different types of testing grips over here.
You will also have this load cell and or change in the elongation connected to a electronics,
which will make the measurement and then give the data output to the computer. So,
you need a control unit and interface. So, this interface would also be basically your
data acquisition unit. Again, a good system would mean that it will acquire
data at a much faster rate and also transmit the data at a very high rate.
Now, once the data goes into the computer, there you need an interface, software interface,
which will read the data and analyze the data. So, you need a software. These are the most basic
components of a UTM. Apart from that, nowadays, you can also have UTM which do measurements at
high temperature or under some humidity conditions. So, you will also have temperature and
humidity setup. And usually, the setup would encompass or will contain the whole
thing inside it so that your sample is properly inside at particular temperature or particular
humidity condition. And in that setup or in that atmosphere, you will be able to do the measurements.
So, the overall system looks very simple, but you would be surprised at what is the cost of such
units. Like I said, a lot of it depends, good quality data depends on a good quality
load frame, a good quality load cell. Then your grips should also be good quality so that there
is no micro slip. So, you are holding the sample like this and it is holding, the grip is holding
like this. But if the grip is not holding properly, there may be micro slip and because of that,
the strain that you get would be erroneous. And no matter how good quality it is, there will
always be some amount of micro slip, but then a good quality machine will try to minimize this.
So, a basic unit of such equipment would cost anywhere between 30 to 40 lakhs,
a very basic unit, Indian rupees. And if you want some atmosphere or even very high strain rate,
then the overall mechanics of the system, the load frame, they need to be much better quality.
And in that case, you will have, so advanced units can cost easily as much as 3 to 4 crore
Indian rupees. So, you can see that this simple looking device or equipment is not so simple.
It has a lot of cost involved with it and therefore a lot of opportunity for innovation,
so that one can make these systems at a much lower cost. In fact, there is an Indian company,
which has now already been bought up by a US conglomerate. So, there is still a lot of
opportunity of innovation in making a lower cost equipment like this.
So, now we have seen the overall setup, but one thing that is missing here is what is actuating
or causing the movement of this crosshead. Now, overall, there are two different types of
setups, which can cause two different, you can say, origin of energy, which causes the actuation
or activation of, actuation of the crosshead. And these two can, this can be classified into
two categories. One is the electromechanical system versus servohydraulic. So, in this one,
you have basically a motor, so something, it may be a linear motor, which causes the movement,
up and down movement of the sample, the crosshead or it can be a screw driven system. So,
it is a motor, which rotates the screw and because of which the crosshead moves up and down.
Or it can be a electromechanical system, which pushes the piston up and down, electromagnetic
system. On the other hand, in servohydraulic, you have a hydraulic system, which is,
which looks something like this. The shape, the overall functioning of this is what allows a very
large amplification of forces here. So, there is fluid in here and there is a piston on one side,
which applies the forces in the other direction. So, you apply the force in this direction,
F1 area, cross sectional area is A1, here the cross sectional area is A2 and this will cause the
force F2. So, you can clearly see F2 is equal to A2 by A1 times F1. So, this kind of system
allows easy amplification of the forces and that is one thing and here you have the liquid,
easy amplification of forces. So, you can apply very, very large loads in this kind of system.
Second is that since it is liquid, the movement is very easily or quickly transferred. So,
movement is very quickly transferred or I would rather directly write the response is very high.
But then this kind of system is also very expensive. On the other hand,
electromechanical systems are much cheaper, but then their drawbacks are that response is not
so high, not so quick. But the good thing is that it is relatively cheaper or less expensive.
So, these are the two different types of system which actuate the crosshead and cause the
displacement of the or elongation of the samples and which are then measured by the electronics,
the displacement as well as the loads and these are captured or recorded by the computer system.
And then you can, once you have the load versus displacement data, then from there you can obtain
the stress, engineering stress versus engineering strain or true stress, true strain up to the UTS
point and beyond the UTS point if you want the data, then like we said earlier, you will need
to get the actual diameter of the neck region and then you can also translate that data into
true stress, true strain. So, that is the one aspect of the machine.
Another aspect that we need to understand about the machine is that what are the practical
problems that may arise. The practical problems that arise, that can arise in a machine and
because of which so much science and engineering goes into making this machine is that when you
are applying the load, you should also realize that this load is also being applied back onto
the equipment and therefore, the equipment will also deform. So, first let me write,
when load is applied, machine parts also undergo deformation. So, what type of deformation would
it be? Would it be elastic deformation or plastic deformation? If it were a plastic deformation,
then after every tensile test, the machine would have permanently changed a little bit and then
it would not be usable. So, clearly it is not plastic deformation, it is only elastic deformation.
And the next question that may arise is that if it is elastic deformation, why are we concerned?
We are concerned because the data that you are getting also involves or includes the elastic
deformation of the system. So, we will see what we mean. So, for example, let us say the crosshead
is moving or crosshead displacement will call which is being recorded by the system,
let us call it delta Lt. Now, this delta Lt recorded by the machine has contribution from
various sources. One, it has the contribution from the elastic deformation of the
system, it has the elastic plus plastic deformation of the sample, which is what you want to actually
measure, this is the only part that you want to measure. And then apart from these, we also get
what is called as micro slip, like we said earlier, that you are holding the sample,
but they will because of the region where the grip is holding the sample, there will be some
amount of slight amount of slipping and that will also add to the total crosshead displacement data.
So, micro slip at the grip. Now, this micro slip at the grip is something that we can minimize,
but we cannot eliminate. For the time being, we will ignore that part. Let us look at what
is the relation between delta Lt and the delta Lm, which is the displacement of the machine
plus delta Lp, which is the displacement of the system. Now, we know that force is equal to
what is the stiffness. So, if stiffness is the C times delta L. Now, stiffness for the,
what is the stiffness of the machine, let us stiffness of the machine, this is related to the
frame, it will be a function of the frame, how good is the frame, how stiff is the frame,
load cell, because load cell is also part of this frame. Then the clamping system,
wherever you have, you are joining two parts, their stiffness will also come into picture.
So, the machine, overall machine stiffness is actually a function of several components,
and these are just important ones. On the other hand, the stiffness of the sample
is a function of the elastic properties. So, basically the slope of the stress-strain curve,
plus it is also a function of the cross-sectional area of the sample.
And apart from this, whether the, how good the sample is made, the quality of the sample,
for example, it may be flat or not so flat. So, all those would go into determining the
stiffness of the sample. Now, if we use this relation and put it over there,
then we know that the overall stiffness of the system total is equal to the stiffness
of the machine plus the stiffness of the sample. And this is a relation which is used to
partition sample strain from total strain. And based on this relation, we also know that
the crosshead velocity that we apply on the, during the elongation is not the same as the sample
deformation rate. So, the V deformation or the V crosshead is equal to Vm. So, the rate at which
the machine is deforming plus the rate at which the sample is deforming.
So, these two things together clearly tell us that the data that we obtain from the equipment
has to be looked at carefully and understood properly. And if you do not carefully partition
the data, you may get very erroneous data. For example, what you would observe is that strain
without any external method of measurement. So, if you are using only the crosshead for measuring
the strain, so let us, usually the system is called extensometer. So, strain without extensometer
are much higher. So, that we can obtain from our relation over here, delta lt is equal to delta lm
plus delta lp. So, this is the delta lt that you are calculating, but this is what you wanted.
So, this is, this one is much higher than what you wanted. And if the strain is much higher,
it means the stress versus strain slope in the elastic region would be much lower. And therefore,
modulus measured without extensometer are much lower. In fact, in one of the recent
labs that is, that we are doing in our campus, in the mechanical behavior lab,
we measured the elastic modulus for aluminum samples, and it came out to close to 12 to 13
gigapascal, when the modulus of elast, for the aluminum is close to 60 gigapascal,
somewhere between 60 to 65. So, you can see that elastic modulus becomes so much lower. Of course,
this is very poor quality data, if we refine it, we can get much higher modulus, but still it will
be lower than the actual value that you expect, which is close to 60 to 65 megapascal. So, next
is that now that we understand that strains calculated using directly from the crosshead
is not accurate. So, what is the better way, what are better ways to calculate strain?
Whenever you are trying to get good quality data used for, which you can use for publication,
then you have to measure strain using one of these techniques. In fact, some of the papers and
reviewers will not accept the data if it has been, if where the strain has been calculated directly
from extensometer. So, one of the methods is extensometer. So, extensometer is basically,
so device which sticks on to the surface of the sample. So, let me draw roughly our sample.
Now, what the purpose of extensometer is that it will attach, will have a knife type two edges
and then it ties up in one way or the other to the sample so that it remains stuck over here.
And then this data will go into the electronics. Now, when the sample is elongated, and this will
become the gauge length for this. So, now when the elongation takes place, it will measure only
the elongation between the changing the distance between these two blades. And therefore, this will
give you a much more accurate because it is not anyways being affected by the stiffness of the
system or the micro slip. So, everything gets excluded. So, this system is able to exclude all
that additional elongation data. Other method is what is called as strain gauge. So, this is
actually a resistance strip, a very, very small resistance strip. So, you apply it over here
on to the surface. So, it has like this. And again, this, these are the two ends of the
resistance wire. And what it is measuring is that when the length elongates,
the resistance changes. So, this system is actually measuring the change in the resistance.
And therefore, it needs to be calibrated a little bit. But once you have calibrated this is this
gives again gives very accurate data. And both of these systems are able to exclude
strain from the system, I will say elongation from the system and micro slip in from the data.
And therefore, this method is very helpful in giving us very accurate data and to be able to
then measure even sensitive data like elastic modulus. So, this would, this lecture would have
given you idea about the equipment, UTM, universal testing machine, or the tensile testing machine.
And we have seen what are the important components and we have seen what are the
limitations and how to overcome some of those limitations. So, with that,
we will come to an end of this lecture. Thank you.